OriGen Biomedical Dual Lumen Catheters

The OriGen Dual Lumen Catheter is produced in size 12F, 15F and 18F. Each catheter is supplied with a blunt introducer in the venous lumen and a self-sealing vent plug in the arterial lumen.

Technical Description

Blunt Introducer

A blunt introducer is supplied for direct introduction via cutdown. Where a guidewire is not utilized when placing the catheter, a blunt introducer can reduce risk of dissection or vessel puncture from using a sharp, pointed introducer. A percutaneous introduction kit is available separately (see below).

Arterial Vent

A self-sealing vent plug in the arterial line connector purges air when introduced into the vein. No need to clamp arterial line during insertion and add weight to the catheter.

Reduced Recirculation

Supply and return holes are positioned away from each other rather than directly opposed. This design has been shown to significantly reduce recirculation (Reference Abstract) and can minimize effects of tissue aspiration. Outflow holes are arranged to jet blood away from catheter.

Constant diameter tip

Solid, radiopaque polyurethane tip. Constant size from tip to Y; no sleeve over inserted portion to hamper insertion into the vein.

Tip Cross section

2 to 1 area ratio: The large (drainage) lumen has approximately twice the area of the return (arterial)lumen

Packaging

Catheters are supplied sterile and non-pyrogenic. Catheters are available individually, and are singly packed in a Tyvek pouch. Guidewires and introducers originally packaged with each catheter, are now packaged separately.

Percutaneous Introducer kit

Because most users place the catheter by cutdown, the items needed for percutaneous introduction have been moved to a separate package. One kit (Item number 'PCTa')is used for all sizes of the OriGen catheter, and is available where introduction over a guidewire is desired. Each kit contains the necessary sharps, guidewires and dilators to safely introduce the catheter.

A guidewire swapping dilator is included in each kit to facilitate changing from a small guidewire to the larger 0.035" guidewire easier. Percutaneous kits are packed individually, and are sterilized with Ethylene Oxide. To download the PCT Instructions for Use, Click Here.

Product codes

V V12F - 12F dual lumen ECMO catheter
V V15F - 15F dual lumen ECMO catheter
V V18F - 18F dual lumen ECMO catheter

PCTa - Percutaneous introducer kit used for all sizes of OriGen Catheters

Placement Recommendations:

The picture below represents ideal placement of the catheter tip in the right atrium: Extended well into the right atrium so the Arterial return holes are pointed at the Tricuspid Valve. If only it were this easy.

Size Recommendations:

The following size recommendations are from Dr. Giles Peek:

  • 12F cannula: patients 2 to 5 kg
  • 15F cannula: patients 4 to 8 kg
  • 18F cannula: patients 7 to 12 kg

Final selection of the cannula size is dependant on the size of the IJV in each particular patient, along with other physiological factors

Catheter Priming Volume

Below is the priming volume of each lumen of the catheters, in ml.

Size
Arterial Lumen
Venous (drainage) lumen
12F
2.8
3.0
15F
3.0
3.5
18f
4.0
5.2

Links:

Click the words below to go to:

Flow Test Reports

12F Flow test results
15F Flow Test results
18F Flow Test results
Arterial Lumen Flow Test comparison
Venous Lumen Flow Test comparison

Get a copy of the instructions for use online:

Catheter Instructions for Use
Percutaneous Kit Instructions for Use


CAUTION: US Federal law restricts the sale and use of this device by or on the prescription of a physician. Supplied Sterile. For Single use only.

1. K. Rais-Bahrami, et al: Improved oxygenation with reduced recirculation during veno-venous extracorporeal membrane oxygenation: Comparison of two catheters. 16th CNMC ECMO symposium, Keystone CO, March 2000

 

Choose a page to jump to:

News

Distributors

Company Information

Regulatory Status

Contact Us

Americas & Asia
OriGen Biomedical Inc.
4020 S. Industrial Dr., Suite 160
Austin, TX USA 78744-1080
Tel: +1 (512) 474 7278
Fax: +1 (512) 617 1503

Europe & Middle East
OriGen Biomedical GmbH
Drottninggatan 13
252 21 Helsingborg
Sweden
Tel: +46 46 772 433
Fax: +46 42 146 074